WRIST STRAP
1. Thread the split ring onto the silver
lanyard anchor located on the
bottom back of the unit. Attach the
strap to the split ring. The strap has
a break away safety feature.
BW DEFENDER KEYPAD
GENERAL:
All operating features and functions are
activated by the rubber keypad. The
keypad operation is designed to prevent
accidental activation of any function.
The BW DEFENDER will advise you
each time a function is accessed, and
when automatic functions, are taking
place. Press the keys down firmly. Each
button has multiple functions, which are
discussed in the appropriate sections of
this manual.
TROUBLESHOOTING
This section lists some common errors and their solutions. If you are unable to solve
the problem, please contact BW Technologies for assistance.
Problem Possible Cause Solution
Unit will not turn ON. No battery.
Battery is dead.
Unit is damaged or defective.
Install battery.
Replace and/or recharge battery.
Contact BW.
Display doesn't show
CO or H
2
S
measurements.
This is a 2-gas version
BW DEFENDER, which does not
measure toxic gases.
Contact BW for kit to upgrade to
the, 4-gas version (D4-2002).
Unit does not display
TWA measurements.
This is a 2-gas version
BW DEFENDER, which does not
measure toxic gases.
Contact BW for kit to upgrade to
the 4-gas version.
Unit does not enter
alarm.
Alarm points are set incorrectly.
Alarm setpoint(s) may be set to
zero (off).
Unit is in calibration mode.
Reset alarm setpoints to correct
values.
Reset alarm setpoints to correct
values.
Recalibrate unit immediately.
Unit enters alarm
immediately when
turned on.
Sensors are not yet stable.
Battery warning alarm -
battery is weak.
Sampling pump is plugged.
Sensor is missing from unit;
or defective sensor.
Wait for 60 seconds.
Replace/recharge battery.
Remove obstruction.
Replace sensor.
Unit does not respond
to keypad.
Battery is dead.
Unit is performing operations that
require no user input.
Replace and/or recharge battery.
Keypad operation will be
automatically restored at the
appropriate time.
Unit does not
accurately measure
gas.
Unit requires recalibration.
Unit is colder/hotter than gas
being measured.
Sensor screens are clogged and
dirty.
Sensor(s) require(s) replacement.
Sensor is missing from unit.
Recalibrate unit.
Allow unit to acquire ambient
temperature before use.
Clean sensors.
Replace sensor(s).
Replace sensor.
Unit does not display
maximum level of
oxygen detected.
Peak detection for oxygen:
displays the lowest value, not the
highest value detected.
Unit is operating normally.
29 6