GETTING STARTED POWER
GENERAL: Familiarize yourself with all operating features and functions of the
BW DEFENDER. The BW DEFENDER is calibrated at the factory. User selectable
Alarm Setpoints have been set to relevant safety standards. The BW DEFENDER will
advise the current Alarm Setpoints each time it is turned ON. If the user wishes to
reset the Alarm Setpoints, refer to “ADJUSTING ALARM SETPOINTS”.
BATTERY AND CHARGER: Batteries must be charged before use.
WARNING: ONLY CHARGE BATTERIES USING A BLACK & DECKER
VERSAPAK™ CHARGER. DO NOT USE ANY OTHER CHARGER.
FAILURE TO OBSERVE THIS PRECAUTION COULD LEAD TO FIRE
OR EXPLOSION.
CAUTION: READ AND OBSERVE ALL INSTRUCTIONS AND PRECAUTIONS IN
THE BLACK & DECKER LITERATURE PROVIDED WITH THE
BW DEFENDER. FAILURE TO DO SO MAY RESULT IN FIRE,
ELECTRIC SHOCK OR OTHER FORMS OF PERSONAL INJURY OR
PROPERTY DAMAGE.
FIRST TIME CHARGE:
The charger supplied with the BW DEFENDER is a VAC
2-port charger, which requires approximately 6 hours (both batteries) for
a first time charge.
NORMAL CHARGE TIME: After the first use, battery charging times decrease.
Normal charge time:
2-port VAC Charger - 3-4 hours for 1 to 2 batteries
Black & Decker VERSAPAK™ chargers (other than the 2-port charger included with
the BW DEFENDER) may be used with BW DEFENDER batteries. Refer to the
instruction manual provided with the specific charger unit for charging instructions.
NOTE: Batteries can be left in the VAC charger until required for use.
OPERATING SPECIFICATIONS
Operating Humidity: 15 to 90%
Operating Pressure Range: Atmospheric ±10%
Note 1: Response typically less than 2 seconds @ T
50
Note 2: All performance data is based on conditions a 20°C (68°F), 50% RH and
1013 mBar (1 atm)
CAUTION: The BW DEFENDER combustible sensor is highly poison resistant. All
catalytic sensors should be protected from exposure to lead compounds,
silicones, chlorinated hydrocarbons. Certain organic vapors (e.g. leaded
gasoline and halogenated hydrocarbons) may temporarily inhibit the
sensor performance, in most cases it will recover after calibration.
Combustible
EX
Oxygen Carbon
Monoxide
Hydrogen
Sulfide
Primary Sensitivity: %LEL % O
2
CO ppm H
2
S ppm
Measuring Range: 0-100% 0-25% 0-500 ppm 0-100 ppm
Maximum Overload:
30 % 1500 ppm 500 ppm
Detectable Accuracy: 1 % 0.1% 1 ppm 0.5 ppm
Response @ T
50
:
6 (Note 1) 2 7 5
Readout Increments: 1% 0.1% 5 ppm 1 ppm
Long Term Drift: Zero:
(% signal loss/year)
Span:
Nominal
<5%
Nominal
<5%
Nominal
<5%
Nominal
<5%
Sensor Warranty: 2 Years 2 Years 2 Years 2 Years
31 4
Updated English D1218/3 Nov/00 P31. (O2 warranty changes to 2 years)