Black & Decker MULTI-GAS DETECTORS Manuel d'utilisateur Page 38

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 71
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 37
BW Defender2
Calibration and Setting Alarm Setpoints
27
Pass Code Protect: The detector will ask for the pass
code before proceeding to Auto Span and Alarm
Setpoints.
Pass Code Protect Activated
Before setting span the display will
advise Calibration is locked.
Then display requests the Pass Code.
The Set up/down arrow icon lights to prompt entry of the
three-digit pass code. Enter the correct pass code using G
and H keys. Press key to accept the displayed pass code.
If the correct code is entered the detector
automatically proceeds to auto span.
If the pass code is not entered within 8 seconds or the wrong
pass code is entered, the detector advises the code is not
correct.
The detector beeps four times and automatically
returns to normal operation.
Vue de la page 37
1 2 ... 33 34 35 36 37 38 39 40 41 42 43 ... 70 71

Commentaires sur ces manuels

Pas de commentaire