BW Defender2
User Manual
26
Calibration Procedure
To calibrate the detector and set its alarm setpoints, perform
the following procedure.
Start Calibration
To quit at any point after auto zero, press . The detector
retains any saved values, and the audible alarm beeps four
times before the detector returns to normal operation.
1. Press A and simultaneously.
The display shows:
The audible alarm beeps once.
Auto Zero and Oxygen Sensor Calibration
The display flashes “auto zero” while the detector
automatically zeroes the H
2
S, CO and LEL sensors and
calibrates the oxygen sensor.
The audible alarm then beeps twice.
Auto Zero Sensor Fail
Note
Do not apply the calibration gas until the display shows a
flashing gas cylinder; otherwise, the auto zero step will
fail.
If a sensor fails, the display advises error (Err) and skips
the span for the failed sensor(s). Other sensors span
normally. Press to exit. Then restart calibration in an
atmosphere that is clear of the targeted gases. If the
auto zero fails a second time, restart the detector to test
the sensors.