GasAlertMax
Troubleshooting
Table 15. Troubleshooting Tips
Problem Possible Cause Solution
The detector does not enter alarm. Alarm setpoint(s) are set
incorrectly
Alarm setpoint(s) set to zero
Detector is in calibration mode
Reset alarm setpoints
Reset alarm setpoints
Complete the calibration procedure
The detector intermittently enters
alarm without apparent reason.
Ambient gas levels are near alarm
setpoint or the sensor is exposed
to a puff of the target gas
Alarms set incorrectly
Missing or faulty sensor
Detector is operating normally. Use
caution in suspect areas. Check
maximum gas exposure reading.
Reset alarm setpoints
Replace sensor
The detector automatically
deactivates.
Automatic shutdown feature
activated due to weak battery
Replace battery
Unit will not auto zero or calibrate.
O
2
sensor reading is erratic.
O
2
sensor replacement may not
be compatible
Change O
2
sensor
55