GasAlertMax
Calibration and Setting Alarm Setpoints
Setting the Calibration Due Date
After all the alarm setpoints have been set, the LCD displays
the number of days remaining before the next calibration is
due. (The default setting is 180 days.)
Note
If you do not calibrate all of the sensors, the LCD
displays notE: SPAn ALL SEnSorS to rESEt CAL
duE dAtE.
The LCD flashes prompting you to change the calibration
due date between 1 and 365 days.
9. If you want to change the next calibration due date,
press H and G until the LCD displays the new
value.
10. Press C to save the new value.
After the calibration due date is set, the audible alarm beeps
four times and the detector returns to normal operation.
Note
If you do not press any pushbuttons within 10
seconds of entering this setting, the detector
automatically retains the existing value.
Verification
After calibration has been completed and the detector is in
normal operating mode, test the detector using a gas cylinder
other then the one used in the calibration. The gas
concentration should not exceed the sensor's detection
range. Confirm that the LCD displays the expected
concentration.
The catalytic LEL sensor must be checked if a fitted H
2
S
sensor gives an alarm (potential poison gas).