GasAlertMax
Calibration and Setting Alarm Setpoints
Calibration Procedure
To calibrate the detector and set its alarm setpoints, perform
the following procedure.
Note
To quit at any point after the auto zero, press
C to
exit auto span. Then press F to exit setting
alarm setpoints. The detector retains any saved
values and the audible alarm beeps four times
before the detector returns to normal operation.
Start Calibration
1. Activate the detector (if applicable).
2. Press C and H simultaneously. The LCD displays
the following:
The audible alarm beeps once.
Auto Zero and Oxygen Sensor Calibration
3. The LCD displays flashes N while the
detector automatically zeroes the H
2
S, CO, and
LEL sensors and calibrates the O
2
sensor.
The audible alarm then beeps twice.
Auto Zero Sensor Fail
Note
Do not apply the calibration gas until the LCD
displays a flashing gas cylinder; otherwise, the auto
zero will fail.
If a sensor fails, the LCD displays Err (error) and
skips the span for the failed sensor(s). Press
C to
exit auto span. Then press F to exit setting
alarm setpoints. Then restart calibration in an
atmosphere that is clear of the targeted gases. If
the auto zero fails a second time, restart the
detector to test the sensors.